FDA Posts Class I Notice for Tandem Mobi Insulin Pump Software Correction

Pumps on three older software versions can falsely detect a motor fault that stops insulin delivery; Tandem is directing users to update to version 7.9.0.2.

Insulin pump with tubing and an infusion set insertion device
A Medtronic insulin pump with an infusion set. Not the Tandem Mobi pump described in this article. David-i98, public domain, via Wikimedia Commons · Public domain

Key points

  • The FDA posted a notice on April 22 identifying Tandem Diabetes Care's correction of Tandem Mobi insulin pumps on software versions 7.6.0.1, 7.6.0.3 and 7.7.0.1 as the most serious type of recall.
  • According to the notice, a false motor failure can trigger Malfunction 12, which stops insulin delivery and cuts communication with the continuous glucose monitor and the Mobi app.
  • Tandem had reported four serious injuries and no deaths associated with the issue as of November 4, 2025; the FDA classified the recall on November 5, 2025.

The US Food and Drug Administration has posted a notice identifying Tandem Diabetes Care's correction of certain Tandem Mobi insulin pumps as the most serious type of recall, according to the agency's recall page, dated April 22. The action covers Mobi pumps running software versions 7.6.0.1, 7.6.0.3 and 7.7.0.1.

The recall involves correcting devices rather than removing them from use, the FDA said. According to the notice, the recall was classified on November 5, 2025, and Tandem sent affected customers an Urgent Medical Device Correction on October 6, 2025.

A false motor failure

Tandem stated that pumps on the affected software "may incorrectly detect a motor issue, known as a false motor failure," the FDA said. The Mobi pump has a vibration motor that gives tactile feedback for alerts, alarms and malfunctions, and the pump's software monitors the electric current flowing through that motor. If the current is detected as out of range, the pump triggers a malfunction alert known as Malfunction 12.

"If your Tandem Mobi pump shows Malfunction 12, insulin delivery will stop, and the pump becomes non-operational," the notice says. The malfunction also ends communication between the pump and the continuous glucose monitor and the Tandem Mobi mobile app. The FDA said that if not addressed, this could result in hyperglycemia, and that in severe cases a user may require hospitalization or intervention from a medical professional.

"As of November 4, 2025, Tandem Diabetes Care has reported four serious injuries, and no deaths associated with this issue," the agency said.

Software update as the fix

The notice lists seven affected catalog entries, including standard, replacement and Medicare versions of the Mobi pump with Control-IQ and a pharmacy starter kit. Tandem's letter told customers to update to software version 7.9.0.2, which the company offers as a free download through the Mobi mobile app, and to have a backup method of insulin delivery ready. Users who receive a Malfunction 12 alert were told to silence the alarm, contact Tandem and switch to their backup method, the notice says.

MedTech Dive reported on April 22 that more than 17,700 devices were affected, according to an FDA enforcement report. The publication noted that the Mobi, a small tubed pump operated from a smartphone, received FDA clearance in 2023, and that it is one of several diabetes device safety problems flagged by the agency this year, including Insulet's March recall of Omnipod 5 pumps over internal tubing tears.

Sources

  1. Insulin Pump Correction: Tandem Diabetes Care Issues Correction for Tandem Mobi Insulin Pumps, US Food and Drug Administration, April 22, 2026
  2. Tandem recalls Mobi insulin pumps over software malfunction, MedTech Dive, April 22, 2026